Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 209

Advertising
background image

4

Parameters

PC Control / Touch Control

197

4.1.1 Dynamic

equivalence

point

titrations (DET) and monotonic

equivalence point titrations (MET)

For equivalence point titrations (DET, MET) the parameters are identical

except for a few exceptions. Both modes are described together here.
Under Start conditions the parameters for the initial addition of titrant

and for the initial measured value (measured value before start condi-

tions are processed) can be edited:
Start volume (d

)

: Volume which is added before the start of the ti-

tration. Under Dosing rate (d) you can enter the rate at which the

start volume is added.

Pause (d): Waiting period which is to elapse before the start of the

titration after the start conditions have been fulfilled.

Start measured value (d): Measured value at which the initial dos-

ing is stopped. Under Dosing rate (d) you can enter the rate to be

used until the start measured value or the start slope have been

reached.

Start slope (d): Slope (measured value per unit volume) at which

the initial dosing is stopped. Under Dosing rate (d) you can enter

the rate to be used until the start measured value or the start slope

has been reached.

The start conditions are processed in succession: 1. Start volume,

2. Start measured value, 3. Start slope, 4. Pause.
Parameters for the Initial measured value:
Signal drift (d): Drift, i.e. the alteration in the measured value per

minute, which has to be reached before the measured value will be

accepted.

Waiting time min. (d): The measured value is not accepted until

the minimum waiting time has elapsed, even if the measured value

drift has already been reached. The minimum waiting time is only

important for drift-controlled measurements.

Waiting time max. (d): If the signal drift is switched off or has not

yet been reached then the measured value will be accepted after

the maximum waiting time has elapsed.

Under Titration parameters you can edit the parameters for the titra-

tion procedure:
Titration rate (d): Three predefined sets of parameters can be

selected for the titration rate (see Table on p. 199). If individual titra-

tion parameters need to be adapted you should select the setting

User and edit the individual titration parameters under User-

defined Parameters.

Temperature (d): Manually entered titration temperature. If a tem-

perature sensor is connected and the temperature measurement is

set to automatic or continuous under Sensor (see Section 4.1.5)

then the temperature will be measured continuously. The value is

used for temperature compensation in pH measurements.

Advertising