Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 256

Advertising
background image

4.5 Calculations

244

PC Control / Touch Control

The selected variable is entered in the formula with [Insert] or [Back].

With [Cancel] you can exit the list without accepting a variable. The in-

dividual variables are described in Section 4.5.7. The character #

stands for a running number (usually from 1 to 9 or 10); this must be

entered manually. If you want to calculate EP1, for example, then EP is

inserted in the formula and you must enter the 1 in the formula editor.
If the method contains more than one data producing command before

the calculation then the command identification will be given before the

variables (e. g. 1M.TITER for the titrant titer used in the first titration).

Data concerning the titrations, measurements and calibrations have the

identification M for mode; evaluation data have the identification E for

evaluation; calculation data have the identification C for calculation. The

command identification (M, E, C) will be counted up separately for each

command type and does not necessarily correspond with the number

of the command line.
[System variables]

opens the list of system variables:

[Results/Statistics]

opens the list containing the existing results

and the statistics variables:

Advertising