9 measured value display – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 48

Advertising
background image

3.7 System settings

36

PC Control / Touch Control

item Export in the File menu and entering the path and file name with

the extension .txt. When the audit trail is exported its contents will be

deleted. The audit trail window is closed with Close.
To print the audit trail click on the symbol in the symbol bar or

choose the report "Audit trail" in the dialog Print/More

Reports/System reports.
Quick filters can be used on the audit trail entries:

Click on a cell in the table which meets the filter criterion.

In the Filter menu click on Quick filter or on in the symbol

bar. You can use the quick filter in succession for various entries.

To show all the entries again click on Show all in the Filter menu

or on in the symbol bar.

You can update the list with so that new entries are also shown. The

audit trail window is not automatically updated "live". When the audit trail

window is opened the table is updated automatically.
You can view an external Audit trail i.e. from a Touch Control as well.

Therefore load the desired Audit trail file in the Audit trail window under

File/Open. With Touch Control the Audit trail file "log.madt" is created

on the memory card in card slot 2. The entire path with file name is

displayed in the title bar of the Audit trail window. An external Audit trail

cannot be deleted in contrast to the internal Audit trail but the other

functions remain the same.
To show the Audit trail of the running PC Control you have to close the

audit trail window and open it again via View/Audit trail.

Note!

An external Audit trail can only be printed with the print symbol .

When printing via the fixed key [Print] the internal Audit trail will be

printed every time even if the external Audit trail is currently displayed.

In the Titrando Compliance Guide there is a step by step description

of the necessary procedure for complying with FDA Guideline 21 CFR

Part 11.

3.7.9 Measured value display

This setting refers only to the display of the measured values in the

determination sequence (live display) and in manual operation (Manual

Control). The values are always stored with their full resolution.

Open the dialog System settings/Measured value display.

Select Number of decimal places with which pH values and

potentials are to be shown in the measured value display.