4.7 Calibrating pH electrodes (CAL pH) and ISE (CAL Conc)

260

PC Control / Touch Control

(Switch on stirrer) The stirrer will be switched on if Stir solution

during measurement

is activated.

Measure buffer 1 The potential of buffer 1 is measured.

(Switch off stirrer) The stirrer will be switched off if the buffer has

been recognized and if Stir solution during

measurement is activated.

Message showing

which buffer has

been recognized.

A message will be shown with instructions for the

next buffer. When the electrode has been im-

mersed in the second buffer the second meas-

urement is started with [OK].

(Stir and/or

pause before

measurement)

The solution is stirred before the measurement.

Pause will be allowed to elapse before the meas-

urement.

(Switch on stirrer) The stirrer will be switched on if Stir solution

during measurement

is activated.

Measure buffer 2 The potential of buffer 2 is measured.

(Switch off stirrer) The stirrer will be switched off if the buffer has

been recognized and if Stir solution during

measurement is activated.

Calculate calibra-

tion data

The calibration data, i.e. the slope and the elec-

trode zero point pH(0) are calculated, the limits

checked and, if necessary, an appropriate mes-

sage displayed. The calibration data will be added

to the sensor data for the particular sensor.

Optional steps are shown in brackets.