3.20.5 Main dialog "live"

The appearance of the main dialog changes slightly when a

determination has been started.

The sample data can also be entered "live", i.e. while a determination is

being carried out (see Section 3.19.1). In calculations the sample data

which has been entered in the main dialog at the time when the

calculation is carried out (CALC command) is always used.
No device-specific settings can be carried out "live"; this is why the

[System] button is not available. It is also not possible to load or create

methods. The [Load method] button is also not available.
With [Hold] you can halt the determination at any time. The system is

then in the HOLD status (status display in the top right corner). The

button marking then changes to [Continue]. You can then use

[Continue] to continue the sequence.
You can switch to the live display with [Live display], i.e. to showing

the method command which is currently being processed.
With [Control] you can alter the settings for the determination

sequence (see Section 3.17). Several functions are not accessible "live".

The settings for Statistics and Sample data silo cannot be altered.

If you deactivate the Autostart checkbox while a determination is

being carried out then the current determination will be concluded but

no new START will be carried out. The sample number cannot be

edited. If you are working with login (see Section 3.7.7) then you can

log out of the system with [Logout] even when a determination is being

carried out. The logout dialog will then be shown immediately. The

determination continues to run in the background. It can be stopped

with the [STOP] fixed key. All other functions are only accessible again

after a new user has logged in.

Advertising