Dosing outside the control range – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 229

Advertising
background image

4

Parameters

PC Control / Touch Control

217

selected minimum rate, the longer it takes until the control point is

reached for the first time.
The following rule of thumb should be considered in order to obtain

as constant a dosing at the control point as possible:

10

L/min

in

reaction

of

rate

Expected

L/min

in

rate

Min.

μ

=

μ

Parameters for the predefined titration rates for STAT

Rate of reaction

50 μL/min

100 μL/min

500 μL/min

Control range

(Dynamics)

1.000 pH

60.0 mV

1.000 pH

60.0 mV

1.000 pH

60.0 mV

Max. rate

0.25 mL/min

0.75 mL/min

2.00 mL/min

STAT

Min. rate

5.00 μL/min 10.00

μL/min 40.00

μL/min

Control range

Control

point

V/mL

t/s

U/mV

V/mL

Fig. 4.10: Reagent addition and control range for STAT

Dosing outside the control range:
Outside the control range the dosing rate is mainly characterized by the

maximum rate

. The parameters maximum rate and dynamics should be

optimized together in such a way that the titration does not overshoot

too much when reaching the control point. The control range should be

set so that the measured value remains within this range while it is kept

constant. A slightly larger control range should be defined for slow re-

actions (e. g. pH = 3, U = 180 mV). A SET pretitration is frequently

used to reach the control point before the substrate is added. This

means that the STAT titration can be started with a small variation from

the control point.