5 Troubleshooting – Maintenance

PC Control / Touch Control

301

5.2.8 822 Curve Simulator

The 822 Curve Simulator can be used as a diagnosis tool for checking

the recording of the titration curves and their evaluation. Titration

methods can be carried out and simulated measuring curves recorded

without the actual presence of a measuring solution.
Details about the connection and operation of this instrument are given

in the appropriate Instructions for Use.
Please note that such a simulation requires that a correctly configured

Dosimat or Dosino must be connected. A normal titration method is

started. This means that the exchange or dosing unit should be

emptied or the corresponding buret tip should be placed in a

sufficiently large collection vessel.

Connection

Switch off the Touch Control or pull out the Titrando mains cable or

the USB plug at the PC (PC Control).

First connect the analog output of the curve simulator with the

electrode input "Ind." of the Titrando. Use the 6.2116.020 Cable

with F-plugs at both ends.

The 822 Curve Simulator is connected to the MSB port of the

Titrando used. Take care that the dosing device is always the last

device behind the curve simulator in as short a chain of MSB

devices as possible (e. g. Titrando-MSB – Stirrer – 822 Curve

Simulator – dosing device).

If the internal dosing drive of the Titrando is used then the 822

Curve Simulator has to be connected to MSB port 2 of the Titrando.

Switch on the Touch Control or start the PC Control and activate the

checkbox "Send dosing signals to MSB2" under

System/Diagnosis/822 Curve Simulator. Confirm the

modification with [Set]. This copies the dosing signals to MSB port

2. Please note that this status can only be deactivated by a system

restart.

Switch on the Touch Control again or reinsert the mains plug and

start the PC Control software.

Simulation of a titration

For the following example of a simulation you should load the

Tutorial method from the examples of methods stored under

Card 1/Examples.

Rotate the curve selection switch on the 822 Curve Simulator to

curve 0 (Titration of an acid with a base).

Start the titration with [Start].

When the titration has been carried out you will obtain the following

example of a curve under Results/Curve: