4.1 Titrations

224

PC Control / Touch Control

At the start of the titration the validity of the titer and the expiry date

of the selected titrant will be checked as well as the GLP test inter-

val for the exchange or dosing unit. The titer (Variable TITER) and

the concentration (Variable CONC) are also available for calculation

(CALC command, see Section 4.5.1).

STAT

, ADD, DOS only: A second dosing device can be defined under

Tandem dosing

in order to enable continuous dosing. Dosing is carried

out with a combination of two dosing devices so that the second dos-

ing device is dosing while the first one is being filled and vice versa.
Dosing device: Select the second dosing device.
Titrant: See above. The concentration, the validity of the titer and

the working life are ignored in contrast to the titrant of the first dos-

ing device.

Filling rate: Rate at which the dosing cylinder of the second

dosing device is filled. The maximum filling rate depends on the cyl-

inder volume of the exchange or dosing unit used (see Sec-

tion 3.8.6

).

However, in order to ensure continuous dosing, the following points

must be taken into account:
• Use the highest possible filling rate in order to keep the filling times

as short as possible. Please note the viscosity and density of the

liquid.

• When you use two dosing devices with different cylinder volumes

the filling rate of the larger cylinder must be at the minimum:

1

.

cyl

.

Dos

2

.

cyl

.

Dos

V

V

1

rate

Filling

2

rate

illing

F

Example:

Dosing device 1: Volume = 20 mL, Filling rate = 50 mL/min

Dosing device 2: Volume = 50 mL
Filling rate 2 ≥ 50 mL/min ⋅ 50 mL/20 mL ≥ 125 mL/min
• The dosing rate must not exceed 75 % of the filling rate of the

smaller cylinder. This corresponds to the following values (at the

maximum filling rate):