3.24 Manual control

186

PC Control / Touch Control

("+" means counterclockwise and "–" means clockwise when seen

from above).

You can carry out manual dosing with [Dose]. The dosing cylinder

will be refilled automatically when a cylinder volume has been

added. Dosing will be carried out for as long as long as the [Dose]

button is pressed down. The addition is always dynamic, i.e.

dosing will be carried out quicker and quicker until the maximum

dosing rate has been reached (starting with 1 mL/min and then

doubling the dosing rate every 1.5 s).

Read the added volume.

You can refill the dosing cylinder with [Fill]. The shown volume

will be reset to 0.0000 mL.

If you exit the dialog with [Back] then the dosing cylinder will be refilled

automatically and the valve switched to the exchange position.

3.24.5 Remote

At the remote box you can set output lines manually. The status of the

input lines is scanned automatically.

Select the [Remote] function in the Manual control dialog.

The current status of the input lines will be shown for each connected

remote box. The output lines can be set. The input and output lines are

always numbered from right to left starting with 0. Consequently, the

output line 0 is set with the signal "00000000000001". Please consult the

online help for entering the binary pattern. You can select binary

patterns defined in the selection list under System/Templates.

The binary patterns defined under Outputs are set with [Set].