3.19 Sample data

148

PC Control / Touch Control

You should now enter the sample data, irrespective of whether you are

using PC Control or Touch Control.

Under Method you can select the name of the method with which

the sample is to be processed. You can use Sample data

silo/Properties to define the Method memory (internal memory,

Card 1, Card 2, Shared memory) from which the method is to be

loaded. You can also enter the method name. If no method is

entered then the current method will be used. If you enter the same

method in successive silo lines then the method will not be loaded

separately for each determination.

If you have switched on the use of the Sample assignment table

under System/Templates/Sample data then you cannot define a

method in the sample data silo. If a method has already been entered

before the sample assignment table is activated then this entry will be

ignored.

You can enter sample-specific data under Identification 1 and

Identification 2.

The sample identifications can be accounted as the variables CI1 and

CI2 (see Section 4.5.7). The labels of the sample identifications can be

edited specifically for the method (see p. 150). Under

System/Templates/Sample data you can create a sample

identification list. You can then make a selection from this template

when entering the identifications. This simplifies the input of frequently

used texts for the identification. You can also enter the identifications

with the aid of a barcode reader or send them from a connected

balance. With Touch Control we recommend the connection of a PC

keyboard if long texts are to be entered as identifications.

Enter the value and unit for the sample size. If you have retained

the standard settings under Sample data silo/Properties then

the silo line will be concluded with the sample size. With PC Control

if you confirm the input of the value with [Enter] then a new silo

line will be generated automatically. With Touch Control if you

confirm the input of the value with [OK] then the program will

automatically change to the edit dialog for the next silo line.

The value for the sample size can be calculated as the variable C00

(see Section 4.5.7). The sample size unit can either be selected from a