1 properties of the result silo – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 176

Advertising
background image

3.22 Result silo

164

PC Control / Touch Control

3.22.1 Properties of the result silo

Under Result silo/Properties you can choose the criteria according

to which the list is to be sorted and the data to be shown for each

determination.

Open the dialog with [Properties].

Select the Contents of Column 1 and Column 2. In Column 1 you

can display the sample data (Identification 1 or 2, or sample size),

determination time, method, the name of the result or the user. In

Column 2 you can select the result to be shown. The numbering of

the results corresponds to the order in which the results have been

calculated in the determination sequence and stored in the result

silo. The numbering does not correspond to the result variable to

which the result has been assigned.

Use Sort by to select the criterion according to which you want to

sort the content of the list. The entries can be sorted in ascending

or descending order.

The maximum number of lines in the result silo can be defined

under Max. number of lines. If a result silo containing more lines

than defined here is imported (loading or backup), the maximum

number of lines is increased to that value.

With [Delete automatically], you define when the contents of the

result silo is deleted automatically. You can set it so that the oldest silo

line will be deleted when the result silo is full and a new determination is

to be entered. All lines can be deleted automatically when the titration

system is switched on, when a new sample series is started with the

autostart function switched on or when the result silo is saved.
You can use filters for the display of the result silo table. A filter is a rule

defining what is to be shown.

Open the dialog for selecting the filter criteria under Result

silo/Properties with [Filter].

Advertising