4

Parameters

PC Control / Touch Control

269

Possible errors of the electrode test

Test criterion

Message

Measure

pH 9:

The drift in stirred solution is

> 1 mV.

General problem

-

Connect electrode.

-

Replace defective elec-

trode cable.

-

Clean diaphragm (see

electrode leaflet).

-

Replace electrode.

pH 9:

–10 mV < U(unstirred)

< 10 mV

and

the sum of drift values after

1, 2, 3 und 4 min is

< 12 mV/min.

Short circuit

Replace electrode.

A response time does not

meet the limit value for the

rating "Usable electrode".

Glass membrane / Dia-

phragm

Clean diaphragm (see elec-

trode leaflet).

The pH values are not de-

fined for all the buffers at the

measured temperature.

No buffer data

Repeat electrode test at a

temperature at which the pH

values of all the buffers are

defined.

All the slopes meet the limit

values for the rating "Usable

electrode"

and

Uoff is outside the given lim-

its.

Unsuitable reference elec-

trode

-

Repeat electrode test with

suitable reference system.

-

Replace contaminated ref-

erence electrolyte.

-

Select electrode type =

Custom and adjust limits

for Uoff.

The streaming potential is

too high.

Contaminated diaphragm

Clean diaphragm (see elec-

trode leaflet).

Two slopes do not meet the

limit values for the rating

"Usable electrode".

Wrong buffer

Repeat electrode test with

buffers pH 4, 7 und 9.

All the slopes do not meet

the limit values for the rating

"Usable electrode".

Partial short circuit

Check temperature sensor

or enter correct calibration

temperature.