4

Parameters

PC Control / Touch Control

251

Variable Description

Method command

MMP

Number of measuring points in the list of

measuring points

DET, MET, SET,

KFT, STAT, MEAS,

DOS

MTS

Stop type (termination of the titration,

measurement or calibration), format: text

DET, MET, SET,

KFT, STAT, MEAS,

DOS, CAL

MEN

Electrode zero point pH(0) or E(0)

DET, MET, SET,

KFT, STAT, MEAS,

DOS, CAL

MSL

Slope of electrode

DET, MET, SET,

KFT, STAT, MEAS,

DOS, CAL

Sensor

MVA

Variance of electrode (mathematically only

defined for three standards or more);

CAL Conc: calculated only if calibration is

carried out with a minimum of four stan-

dards

MEAS Conc, CAL

Conc

FP#

Fixed endpoint volume for FP#

EVAL FIX-EP

FM#

Fixed endpoint measured value for FP#

EVAL FIX-EP

FT#

Fixed endpoint temperature for FP#

EVAL FIX-EP

Evaluation

fixed endpoints

FD#

Fixed endpoint time for FP#

EVAL FIX-EP

HP#

pK/HNP volume for HM#

EVAL pK/HNP

HM#

PK value, or half neutralization potential

EVAL pK/HNP

HT#

pK/HNP temperature for HM#

EVAL pK/HNP

Evaluation

pK value / half

neutralization

potential

HD#

pK/HNP time for HM#

EVAL pK/HNP

XIP

Volume to minimum measured value

EVAL MIN/MAX

XIM

Minimum measured value

EVAL MIN/MAX

XIT

Temperature to minimum measured value

EVAL MIN/MAX

Evaluation

minimum

XID

Time to minimum measured value

EVAL MIN/MAX

XAP

Volume to maximum measured value

EVAL MIN/MAX

XAM

Maximum measured value

EVAL MIN/MAX

XAT

Temperature to maximum measured value EVAL MIN/MAX

Evaluation

maximum

XAD

Time to maximum measured value

EVAL MIN/MAX

BP#

Breakpoint volume for BP#

EVAL BREAK

BM#

Breakpoint measured value for BP#

EVAL BREAK

BT#

Breakpoint temperature for BP#

EVAL BREAK

Evaluation

break point

BD#

Breakpoint time for BP#

EVAL BREAK