Usb sample processor, Tower 1 and 2 – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 72

Advertising
background image

3.10 Device manager

60

PC Control / Touch Control

USB Sample Processor
The Program version, the Serial number and the name of the

current sample rack are shown for the USB Sample Processor

connected.
With [Adjustment data] you can access an overview of the EEPROM

data of the USB Sample Processor.

Attention!

Do not modify these settings under any circumstances. This data is

required by the service technician in case of positioning problems.

Tower 1 and 2

Note!

For a USB Sample Processor with two towers

Tower 1 is the right-

hand tower when seen from the front and Tower 2 the left-hand

one.

• Under Maximum stroke path you can enter the lowest permissible

lift position for automatic and manual operation. This setting of the

max. stroke path is important for safety reasons. If this value is set

correctly then electrode breakage can be avoided, as a titration

Advertising