Usb sample processor, Tower 1 and 2 – Metrohm 840 PC Control 5.0 / Touch Control User Manual
Page 72
3.10 Device manager
60
PC Control / Touch Control
USB Sample Processor
The Program version, the Serial number and the name of the
current sample rack are shown for the USB Sample Processor
connected.
With [Adjustment data] you can access an overview of the EEPROM
data of the USB Sample Processor.
Attention!
Do not modify these settings under any circumstances. This data is
required by the service technician in case of positioning problems.
Tower 1 and 2
Note!
For a USB Sample Processor with two towers
Tower 1 is the right-
hand tower when seen from the front and Tower 2 the left-hand
one.
• Under Maximum stroke path you can enter the lowest permissible
lift position for automatic and manual operation. This setting of the
max. stroke path is important for safety reasons. If this value is set
correctly then electrode breakage can be avoided, as a titration