Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 156

Advertising
background image

3.19 Sample data

144

PC Control / Touch Control

is shown in the main dialog, but can be edited. A fixed sample size

cannot be entered in the main dialog. You can also define sample size

limits in the method which will be monitored during the determination

sequence (START of the determination and at the end of the

determination) (see Section 3.16.6).

Note!

The

sample size limits are not monitored during the input of the

sample data in the main dialog.

3.19.2 Sample data request in the determination sequence

To make sure that the sample data input is not forgotten you can also

request the sample data in the determination sequence with a

REQUEST command. The automatic sample data request in the

determination sequence is essential for back-weighing. In the method

you can define which data (Identification 1 and/or 2, sample size) is to

be requested (see Section 4.13.3).

If the Hold sequence function is activated in a REQUEST command

then the determination sequence will be stopped and must be

continued when the sample data has been entered with [Continue]. If

the Hold sequence function is deactivated then the determination will

continue in the background until the measured data is required for

further processing. The determination will then be holded automatically

and only continued when the sample data has been entered. In this way

you can ensure that the sample data required for calculations carried

out after the titration, measurement etc. is actually available. In this case

the sample data input is also confirmed with [Continue].
If sample size limits have been defined in the method then these will

be monitored when the entry has been confirmed with [Continue] (see

Section 3.16.6). If a fixed sample size has been defined in the method

then the sample size will be shown during the input of the sample data

but cannot be edited (see Section 3.16.6).