3

Operation

PC Control / Touch Control

115

3.16.2 Insert command

A method may contain a max. of 99 commands. You can use [Insert

command] to insert a new method command in front of the selected line.

In the example shown a new "empty" method is being edited.

Use [Insert command] to open the dialog Sequence/Insert

command.

The method commands are arranged in different groups. An overview

of all commands and parameters can be found in Section 4. The

individual commands are also described in the online help.

Select the group from which you wish to insert a command. For

Report the command is inserted directly. For all other groups the

command selection list is opened for you to select a command.

After a command has been inserted the list of commands is

automatically shown again. The inserted command is shown as

selected so that you can use [Edit command] directly to edit the

parameters of the newly inserted command (see Section 3.16.1). You

can insert a maximum of nine titrations or measurements and nine

calculations. There is no limit to the number of other commands (up to

99 in one method).

Advertising