5 calculations, 1 calc command – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 250

Advertising
background image

4.5 Calculations

238

PC Control / Touch Control

4.5 Calculations

In the method sequence result calculations can be defined with the

CALC command. A method can contain a maximum of nine CALC

commands each with nine result calculations. The CALC commands

also contain instructions for statistical calculations, titer assignments,

storing common variables and storing results in the result silo. Limits

can also be defined for the results. A range of calculable variables

(raw data from the determination, previously calculated results and sys-

tem-specific variables) is available for the calculations. Results which

are calculated in the method sequence can be entered as variables for

parameters.
During titrating or measuring a calculated result can be displayed live

with the CALC LIVE command. The current volume or measured value

respectively is calculated for this purpose. This is helpful e. g. for KF ti-

trations in order to follow the water content during the titration.

4.5.1 CALC

command

Up to nine result calculations can be defined in a CALC command. The

results will be calculated in the sequence which is defined by the num-

bering of the result variables: R1...R9.

In the list of results you can use [Edit] to enter the equations and pa-

rameters for the result calculation and delete the selected result with

[Delete]

.

You can insert a new result calculation in the list with [New]. The list

with Metrohm result templates is opened.

Advertising