Operation

PC Control / Touch Control

39

In the list of titrants the name, cylinder volume and type, i.e.

exchange unit (EU) or dosing unit (DU) are given for every titrant.

Intelligent exchange or dosing units are indicated by IEU and IDU and

shown in green color (magenta if the line has been selected). For

intelligent exchange or dosing units that are attached, the dosing

device connection is shown additionally and – if several control

devices are configured – the name of the control device appears as

well.
If data is read from the data chip of an intelligent exchange or dosing

unit then a check is made whether the titrant list already contains a

titrant in the same type of exchange or dosing unit with the same serial

number. If this is the case then the older data record will always be

overwritten by the new data record, no matter whether the data record

in the titrant list of the system or on the data chip is the most recent

one.
You can configure a new titrant with [New] (see Section 3.8.1 and

Section 3.8.2). With [Delete] titrants can be removed from the list.

Titrant data for intelligent exchange or dosing units can be deleted even

when these have just been attached. The data will be entered in the list

again the next time that the corresponding unit is attached and the data

read out. [Edit] opens the dialog for entering titrant data.

3.8.1 Configuring a titrant in a new intelligent exchange or dosing

unit

Attach the new exchange unit to the Titrando or Dosimat or attach

the new dosing unit to the Dosino (see manual for the Titrando or

for the Dosino).

When the exchange or dosing unit has been recognized the following

message will appear: