INFICON HAPSITE ER Chemical Identification System User Manual

Page 112

Advertising
background image

3 - 28

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

8

By touching CMPD ID, a list of found compounds will appear. See

Figure 3-47

.

This page will display for each compound:

The CAS number

The Fit

The retention time

TIC (Total Ion Count) Max,

The current TIC

The Run Time which is the time remaining until the run is finished

NOTE: Touching a compound on the list will display its Synonym and Exposure

Limit information if it is contained in the NIOSH database.

8a

The CMPD ID screen can also be accessed by using the arrow keys to
highlight CMPD ID and pushing OK SEL.See

Figure 3-47

.

Figure 3-47 Sample Compound Identification for Concentrator

9

To view the chromatogram while the method is running, touch GRAPH. See

Figure 3-48

. Alternately, use the arrow keys to highlight GRAPH. Push OK

SEL. This screen will display:

The TIC Max

The current TIC

The time remaining until the run is finished

NOTE: Touching the blue compound identification above the chromatogram

will display its NIOSH database information. See

Section 3.6, The Help

Icon, on page 3-41

.

Advertising