2 placing the hapsite on the service module, Caution – INFICON HAPSITE ER Chemical Identification System User Manual
Page 59
2 - 21
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
Figure 2-24 Cleaning the Service Module Rubber O-ring using a Lint-free Wipe
After cleaning the rubber o-ring, make sure there are no cuts on it. Also, look for
any visible cracking of the o-ring. If the o-ring is damaged, the o-ring may need to
be replaced. Ensure the o-ring is completely seated in the groove.
2.6.2 Placing the HAPSITE on the Service Module
CAUTION
Never operate the HAPSITE on the Service Module in a
moving vehicle.
1
Remove the green plastic protective cover from the bottom of the HAPSITE.
The green cover is shown in
Figure 2-25
.
NOTE: Store the yellow cover in a clean, dry place, where it will be easily
accessible for later use. This protective cover keeps dust and debris
out of the HAPSITE manifold connection.
CAUTION
During decontamination of the HAPSITE, the yellow
protective cover must be installed.