HAPSITE ER Operating Manual

2a

If the Actual Percentage is above the acceptable range, (red box appears to
the right of the column, indicating the mass response is too high) the ion energy
needs to be reduced. Lower the Ion Energy by increments of 5 at a time.

2b

If the Actual Percentage is below the acceptable range, (red box appears to
the left of the column, indicating the mass response is too low) the Ion Energy
needs to be increased. Raise the Ion Energy by increments of 5 at a time.

Figure 7-35 Ion Energy Needs to be Increased

NOTE: When adjusting Ion Energy, there are low and high limits for each

mass. Make sure manual adjustments stay within the limits shown in

Table 7-1

.

Table 7-1 Manual Adjustments Guidelines

MASS

IE low

IE high

50 90 170

55 90

170

69

90 170

93

90

170

117 140 170

167

140

220

213 175 230