HAPSITE ER Operating Manual

16.5.4 Decontaminating the Outside of the Analytical Module

1

Decontamination cap and plug kit number 930-2020-G1 contains tall the plugs
and caps needed to seal the Analytical Module for decontamination. See

Figure

16-3

.

Figure 16-3 Plugs and Caps

2

Install the 24 V Power electrical cap (P/N 070-1273). See

Figure 16-4

.

3

Install the EXT COM electrical cap (P/N 070-1273). See

Figure 16-4

.

Figure 16-4 24 V and EXT COM Caps