HAPSITE ER Operating Manual
CAUTION
All unknown samples should first be collected with the
Survey Method to avoid saturating the GC. The Survey
Method allows the operator to gain knowledge of how far
away to hold the probe from the sample and if a
Concentrator needs to be used.
The sampling procedure for an Analyze (GC/MS) method is dependent on the
HAPSITE configuration. For information on running Concentrator, HSS and/ or
SituProbe samples, refer to either
for information on running them from
the Front Panel or
for information on running them from the Laptop.
6.2.1 Building an Analyze (GC/MS) Method and
Target Compound Library File
Collection of Analyze (GC/MS) data is controlled by the method. Analyze (GC/MS)
methods are generated or modified using the Method Editor. The Method Editor is
covered in detail in
. The Target Compound Library is
built using the Calibrate Function. The Calibrate Function is described in detail in
Chapter 12, Target Compound Methods
.