HAPSITE ER Operating Manual

15.9 Physically Removing the HAPSITE from the Service Module

The HAPSITE cannot be physically removed from the SM until a "Detach SM" has
been performed. Refer to

section 15.8, Detaching the HAPSITE, on page 15-16

before physically removing the HAPSITE from the Service Module.

After the HAPSITE has been detached from the Service Module, release the
latches on each side of the Service Module.

HINT: To continue using the HAPSITE, place a charged battery in the HAPSITE

or connect exterior power before removing the HAPSITE from the docked
position on the Service Module.

Physically remove the HAPSITE from the Service Module. Lift the HAPSITE off the
Service Module in a straight, upward direction. Once the HAPSITE is undocked,
replace the yellow plastic protective cover on the bottom of the HAPSITE, as
shown in

Figure 15-28

.

Figure 15-28 Placing Yellow Protective Cover on Bottom of HAPSITE

This yellow cover is used to keep out dust and debris, as well as keep the opening
dry during decontamination.