INFICON HAPSITE ER Chemical Identification System User Manual

Page 413

Advertising
background image

12 - 17

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

Figure 12-17 Save Library Button

12

The template is complete and the library is ready for Calibration. Refer to

Figure

12-18

.

Figure 12-18 Library Ready for Calibration

12a

Prior to calibration, the curve fit should be selected. To change the curve fit from
the default Linear, Forced through Origin, select the View/Edit button under
the Method controls. Select the curve fit and save the Method.

12b

To calibrate the library, select the standard Data Files, select Calibrate Library
and press the Start button under Peak Search.

12c

Review the curves for each analyte. To review each analyte, select the
Calibration tab and use the mouse to select each compound. See

Figure

12-19

.

Figure 12-19 Calibration Curve