HAPSITE ER Operating Manual
Figure 12-17 Save Library Button
12
The template is complete and the library is ready for Calibration. Refer to
.
Figure 12-18 Library Ready for Calibration
12a
Prior to calibration, the curve fit should be selected. To change the curve fit from
the default Linear, Forced through Origin, select the View/Edit button under
the Method controls. Select the curve fit and save the Method.
12b
To calibrate the library, select the standard Data Files, select Calibrate Library
and press the Start button under Peak Search.
12c
Review the curves for each analyte. To review each analyte, select the
Calibration tab and use the mouse to select each compound. See
.
Figure 12-19 Calibration Curve