HAPSITE ER Operating Manual

7.2 AutoTune

There are two versions of the AutoTune program: Short and Long.

The Short AutoTune program is the daily maintenance tune of the HAPSITE. It is
pre-configured to run automatically at startup, after the system has been powered
OFF and ON with a method selected, or after an twelve hour period of operation.
Short AutoTune can also be invoked from the Manual Tune function at anytime.
The Short AutoTune will check if all parameters are within specifications, record the
dwell statistics (Baseline/Threshold), create a Tune report and proceed to the Run
Method function. If corrections are required, the program will automatically make
the correction, measure the dwell statistics, create the report, and then proceed to
the Run Method function.

and

Figure 7-1

shows the laptop screen during a Short AutoTune. The status

screens will vary depending on the corrections or adjustments required.

Figure 7-1 Status Screen from Laptop During Short AutoTune