2 performance specifications, 3 serial number location – INFICON HAPSITE ER Chemical Identification System User Manual

Page 30

Advertising
background image

1 - 2

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

1.2 Performance Specifications

The performance specifications for the HAPSITE are shown below:
Mass range . . . . . . . . . . . . . . . . . . 1-300 AMU
Scan Rate . . . . . . . . . . . . . . . . . . . as much as 1000 AMU/sec @

10 points per AMU

Ionization Mode . . . . . . . . . . . . . . . 70 eV electron impact
Operating temperature range . . . . . 0

o

C to 45

o

C (32 to 113

o

F)

Dimensions (LxWxH) . . . . . . . . . . . 46 x 43 x 18 (cm); 18 x 17 x 7 (in)
Weight . . . . . . . . . . . . . . . . . . . . . . 16 kg (35 lb.) without the battery
Internal Power Consumption . . . . . 30 watts average, 24 V(dc)
Carrier Gas . . . . . . . . . . . . . . . . . . Nitrogen
Column Temperature Range . . . . . 60

o

C to 200

o

C

Maximum Sample
Moisture Content . . . . . . . . . . . . . . . 8% by weight
pH Range of Sample . . . . . . . . . . . . 2 - 11
Boiling Point of Sample (approx) . . . <2

70

o

C

Chemical Composition of Sample . . 1 - 1

5 Carbon atoms

GC Column

. . . . . . . . . . . . . . . . . . . 100% dimethyl polysiloxane phase, 15 m X 0.25

mm ID

x 1.0

m film

SIM Channels . . . . . . . . . . . . . . . . 10
External Communications . . . . . . . Ethernet Port, Wireless
Carrier Gas Use-Rate . . . . . . . . . . One 5L canister per 8 hours of operation

(This depends on the details of the method
being used.)

Internal Standards Gas Use-Rate . One 5L canister per 24 hours of operation

(This depends on the details of the method
being used.)

Battery Life . . . . . . . . . . . . . . . . . . Approximately 2 to 3 hours

1.3 Serial Number Location

The serial number of the HAPSITE is located on the inside of the front panel and
in the front panel software. Touch the HAPSITE icon, followed by the HAPSITE
System
icon, and then touch HAPS to locate the serial number.