HAPSITE ER Operating Manual

3

Select default.tun file and press OK. See

Figure 7-29

.

Figure 7-29 Select default.tun File

4

Wait for the automated process of opening the tune to complete.

5

Check the Base Peak Gain (BPG). Is the value between 0.4 and 0.6? The
optimal setting for the BPG is 0.5. See

Figure 7-30

and

Figure 7-31

.