INFICON HAPSITE ER Chemical Identification System User Manual
Page 252
Advertising
7 - 26
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
3
Select default.tun file and press OK. See
Figure 7-29 Select default.tun File
4
Wait for the automated process of opening the tune to complete.
5
Check the Base Peak Gain (BPG). Is the value between 0.4 and 0.6? The
optimal setting for the BPG is 0.5. See
and
Figure 7-31
.