HAPSITE ER Operating Manual

7.6 Performing Manual Tune

CAUTION

Manual Tune is a procedure that should only be
attempted by experienced HAPSITE users who have
taken the Advanced Training course provided by
INFICON.

Read

section 7.5, Manual Tune Settings and Controls, on page 7-9

to learn about

the various controls and settings involved in Manual Tune. If the system will not
tune after repetitive tuning, call INFICON Support for additional guidance. Below
are general guidelines for performing a Manual Tune.

NOTE: Do not save a Tune file from a Manual Tune if it indicates the HAPSITE is

not tuned.

7.6.1 Adjusting Base Peak Gain

1

Double-click on the ER IQ icon. See

Figure 7-27

.

Figure 7-27 ER IQ Icon

2

Double-click on the Tune icon. See

Figure 7-28

.

HINT: Manual Tune can only be opened in Advanced User Mode. See

section

8.8.1, Changing Access Levels, on page 8-18

for instructions on

changing the access level.

Figure 7-28 Manual Tune Icon