3.1.4 Quick Reference SOP - Heat-up and Tune
CAUTION
Do not open the front panel in a wet or contaminated area.
1
Insert the internal standard and carrier gas canisters.
2
Insert a charged battery.
3
Connect the AC to DC power converter power supply.
4
Verify that the appropriate sample configuration (i.e., concentrator) is installed.
5
Press the POWER button on the front panel.
6
The HAPSITE will heat up the necessary components and perform AutoTune.
A prompt to run SURVEY or ANALYZE will appear when the HAPSITE is ready
to run a sample. The process for heating and tuning takes approximately 20
minutes.
7
If the default method is not the desired method, touch STOP PREPARE.
8
Touch SELECT METHOD. Highlight the desired method. Touch Select.
NOTE: If the wireless connection to the Laptop is to be used, see
Chapter 4,
Communications and Touch Screen Options
.
NOTE: When the SYSTEM READY message is displayed, touch either RUN
SURVEY or RUN ANALYZE. If using the push buttons, push SURVEY
RUN or ANALYZE RUN.