1 headspace – INFICON HAPSITE ER Chemical Identification System User Manual

Page 350

Advertising
background image

11 - 12

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

11.4.1 Headspace

CAUTION

Only trained users should create methods. Choosing the
incorrect sampling device and/ or temperatures may
result in inaccurate data.

To create a Headspace method, choose the Headspace Sample Input Device.
This will bring up another temperature setting window for setting the Headspace
Oven and Transfer Line temperature. See

Figure 11-10

.

Figure 11-10 Headspace Parameters

Headspace Temperatures in Celsius

Oven . . . . . . . . . . . . . . . . . . . . . . . . The target temperature the Oven must

achieve before starting data acquisition. This
setting is not available when the Headspace
is disabled.

Transfer Line. . . . . . . . . . . . . . . . . . The target temperature the Transfer Line

must achieve before starting data
acquisition. This setting is not available when
the Headspace is disabled.

Advertising