HAPSITE ER Operating Manual

The SKIP HEATER option allows for the HAPSITE to run when the probe is cold.
This is useful in cold climates, when it may be difficult for the probe to heat to the
set point temperature. See

Figure 3-101

.

Figure 3-101 Probe Icon

If the Service Module is attached, the Service Module icon will be shown. Touching
the Service Module icon will display the installed version of the service module’s
firmware and the turbo speed of the service module’s pump. Options for attaching
and detaching the service module can be found through this screen. For more
information on the service module, see

Chapter 3, Operating HAPSITE in Portable

Mode

.

The Headspace Sampling System icon will appear when it is attached to the
HAPSITE. Touching the HSS or SituProbe icon will display the information
regarding the firmware version and accessory’s heaters. See

Section Chapter 13,

Headspace Sampling System, on page 13-1

and

Section Chapter 14, SituProbe,

on page 14-1

for further information.

3.8 EXIT Menu

The EXIT menu is located on the top row of the front panel. See

Figure 3-102

. This

option will access Turn Off, Reboot or Standby. Turn Off will shut down the
HAPSITE’s power. Reboot will reset the microprocessor in the HAPSITE and
reload the drivers. It will also restart the operating system, the HAPSITE ER
program and the front panel program. The Standby option will put the system into
Extended Standby. Refer to

section 3.8.1 on page 3-61

for Extended Standby

instructions.

1

Touch EXIT. Alternately, use the arrow keys to highlight EXIT and push OK
SEL
. See

Figure 3-102

.