2 removing the concentrator – INFICON HAPSITE ER Chemical Identification System User Manual
Page 70

2 - 32
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
NOTE: The Concentrator Cover contains two metal contacts, inspect the
contacts, prior to assembly, to be sure the contacts are not bent or
crimped. See
Figure 2-45
.
Figure 2-45 Back of Concentrator Cover
9
Close the front panel.
2.7.2 Removing the Concentrator
1
Open the front panel of the HAPSITE and remove the black cover labeled
CONCENTRATOR.
WARNING
The elbow fittings may be hot. Allow these components
to cool before continuing.
WARNING
Excessive force and/or tightening can cause the fragile
glass to break!
2
With fingers, loosen the nuts on the top and bottom of the Concentrator until the
it becomes free.
3
Lift the top elbow. Gently lift and angle the concentrator tube out of the fixture.
4
Remove the Concentrator from the bottom elbow, being careful not to lose the
ferrules inside the nuts.
5
Store the Concentrator wrapped in tissue in its storage vial for future use.