3.1.3 Concentrator Cleanout Failure

If the concentrator cleanout is successful, the screen will display the final TIC. See

Figure 3-20

.

Figure 3-20 Cleanout Successful

If the concentrator cleanout is unsuccessful, the screen will display a concentrator
cleanout failed message. See the instructions below for cleanout options.

1

Touch Retry to start another concentrator cleanout sequence.

2

Touch Skip to start running a concentrator Analyze method.

3

Touch Abort to return to the Main Screen.

NOTE: If Abort is touched, the HAPSITE will show that the SYSTEM IS NOT

READY.

4

The HAPSITE will re-run the cleanout as part of its preparation.

5

If the failure box appears again, check the concentrator to verify that it is not
cracked or chipped. Also, try re-installing the concentrator to ensure that it is
properly seated.