3.1.3 Concentrator Cleanout Failure
If the concentrator cleanout is successful, the screen will display the final TIC. See
Figure 3-20
.
Figure 3-20 Cleanout Successful
If the concentrator cleanout is unsuccessful, the screen will display a concentrator
cleanout failed message. See the instructions below for cleanout options.
1
Touch Retry to start another concentrator cleanout sequence.
2
Touch Skip to start running a concentrator Analyze method.
3
Touch Abort to return to the Main Screen.
NOTE: If Abort is touched, the HAPSITE will show that the SYSTEM IS NOT
READY.
4
The HAPSITE will re-run the cleanout as part of its preparation.
5
If the failure box appears again, check the concentrator to verify that it is not
cracked or chipped. Also, try re-installing the concentrator to ensure that it is
properly seated.