16.6.2 Reactivating the NEG Pump
Reactivating the NEG Pump requires having the Service Module attached to the
HAPSITE for at least 22 hours. The procedure is basically the same as activating
a new NEG pump. Refer to
section 2.6, Service Module, on page 2-17
information on setting up the Service Module.
1
Click on the Service Module tab of the Properties menu.
2
Click on the Service button. See
Figure 16-12 Service Button on the Service Module Tab
3
Use the Activate NEG settings below as a guideline. However, change the
HAPSITE Pumpdown Time to 180. See