16.6.2 Reactivating the NEG Pump

Reactivating the NEG Pump requires having the Service Module attached to the
HAPSITE for at least 22 hours. The procedure is basically the same as activating
a new NEG pump. Refer to

section 2.6, Service Module, on page 2-17

for more

information on setting up the Service Module.

1

Click on the Service Module tab of the Properties menu.

2

Click on the Service button. See

Figure 16-12

.

Figure 16-12 Service Button on the Service Module Tab

3

Use the Activate NEG settings below as a guideline. However, change the
HAPSITE Pumpdown Time to 180. See

Figure 16-13

.