HAPSITE ER Operating Manual

Figure 15-26 Detach In Process

5

The HAPSITE Vacuum Interconnect valve will close shortly after selecting
DETACH.

NOTE: The detach procedure typically takes about 3 to 5 minutes to complete.

NOTE: After pressing the Detach button and selecting Yes on the confirm

window, the HAPSITE Properties window can be closed.

6

When the Detach procedure is complete, a message will appear confirming a
successful detachment. See

Figure 15-27

.

Figure 15-27 Detach Successful

NOTE: To place the Service Module in storage see

section 15.10, Storing the

Service Module, on page 15-20

.

15.8.2 Using the Front Panel Display to Detach

1

Touch Menu or push MENU.

2

Touch the PROBES & SM icon. Alternately, push the STAT button until the
PROBES & SM page is displayed.

3

Touch the DETACH SM button. Alternately, using the arrow keys, highlight the
DETACH SM button. Push OK SEL

4

A bar graph displaying the progress of the Detach procedure will appear on the
front panel of the HAPSITE.

5

The message Service Module Found, Not Attached will be displayed on the
screen.

CAUTION

Wait until the Turbo Speed reads zero before physically
removing the HAPSITE from the Service Module.

6

To place the Service Module in storage, see

section 15.10, Storing the Service

Module, on page 15-20

.