HAPSITE ER Operating Manual
CAUTION
Make sure all tools are clean by wiping them with
methanol. Wear powder free gloves for the rest of this
procedure.
10
Insert the forked end of the ionizer removal tool (see
Figure 16-46
) behind the
collar on the stainless steel tube about 1/4 inch from the left side manifold wall.
11
Insure the stainless steel pin on the ionizer removal tool is facing toward the left
manifold wall.
Figure 16-46 Ion Source Removal Tool
12
Apply pressure, using the stainless steel pin as a leverage point. Slide the
extension tube from the membrane isolation valve into the ion volume. The
extension tube should disengage completely from the membrane isolation
valve.
NOTE: If the extension tube does not slide easily into the ion column, loosen
the ionizer. Then use the ionizer removal tool to reposition the ionizer,
which will enable the extension tube to slide into the ion volume as
displayed in
Figure 16-47
.