HAPSITE ER Operating Manual

Figure 11-8 Method Editor Startup Page for Survey Methods

The parameters on the Startup page are:

Use Internal Standard . . . . . . . . . . This option is shown on the Startup Page for

Survey methods. Refer to

Figure 11-8

.

HAPSITE Temperatures in Celsius

Column . . . . . . . . . . . . . . . . . . . . . . The target temperature the Column must

achieve before starting data acquisition. (Not
used for Survey methods.)

Membrane . . . . . . . . . . . . . . . . . . . . The target temperature the Membrane must

achieve before setting before starting data
acquisition.

Valve Oven . . . . . . . . . . . . . . . . . . . The target temperature the Valve Oven must

achieve before starting data acquisition.