HAPSITE ER Operating Manual

6

OPTIONAL: Insert a charged battery into the SituProbe accessory battery slot.
The battery is not required when using the external AC to DC HAPSITE
Adapter.

7

Connect the Nitrogen purge and sample lines on the SituProbe purge head
(IPN 932-200-G1) onto the side of the SituProbe accessory. See

Figure 2-34

.

Figure 2-34 Connecting Purge Head

8

Power on the SituProbe accessory by using the toggle switch located inside the
front panel door. Close the front panel door before using the SituProbe.

9

If using a purge vessel, fill the vessel with 1L of water. Place the SituProbe
purge head into vessel. Sample will flow inward through the side bottom
opening and outward through the top side opening. See

Figure 2-35

.

Figure 2-35 Purge Vessel Water Flow Direction

10

The purge vessel can also be used with the openings capped. In this option, fill
the vessel with 1L of the water to be sampled. Place the purge head into the
vessel.

11

Alternately, place the SituProbe purge head into the water supply to sample.
NOTE: Minimum submersion depth is 5.5 inches (14 cm) and the maximum is

just below the top nut.The depth should be measured from the bottom
of the purge head towards the electrodes.

CAUTION

If the purge head is submerged in a depth greater than
recommended, water in the HAPSITE may result.