INFICON HAPSITE ER Chemical Identification System User Manual

Page 149

Advertising
background image

3 - 65

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

3.9 Analyze (GC/MS) Mode with Headspace Sampling System in

Portable Mode

The Headspace Sampling System is used to test liquid and soil samples.

NOTE: See

Chapter 13, Headspace Sampling System

for more information on the

Headspace System. Refer to

Section 2.5, Headspace Sampling System,

on page 2-12

for assembly instructions.

The Concentrator method is used for analyzing samples with concentration levels
in the part per trillion range. See

Chapter 6, Methods

for additional information on

GC/MS methods.

1

Verify that the desired sampling configuration is installed (i.e., Concentrator).

2

Make sure that the Headspace Transfer Line is installed.

3

Place a 40 mL vial with either blank or sample into a Headspace well.

NOTE: Do not fill 40 mL vial more than 20 mL full. See

Section 13.2.3, Loading

the Wells, on page 13-11

for detailed instructions and information on

loading the sample vials into the wells.

4

Place a blank vial into a Headspace well.

WARNING

The Headspace needle is very sharp and the assembly
may be hot.

Figure 3-112 Headspace Needle

Advertising