4 tune files – INFICON HAPSITE ER Chemical Identification System User Manual
Page 328
10 - 4
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
Figure 10-4 Method Editor - Showing File Name Options
The system allows flexibility to add the Date or the Date and Time as a prefix or
suffix. A different naming convention can be used if Use Default Filename is not
selected.
10.1.4 Tune Files
Refer to
for additional information on Tune.
NOTE: Manual Tuning requires Advanced access. Normal users do not have
access to Manual Tune. Refer to
section 8.8, Access Levels, on page 8-17
for additional information on Access Levels.
If Manual Tune is performed, the file must be saved to the Tune folder on the
HAPSITE.
Figure 10-5
shows a typical Manual Tune screen, which is opened by
selecting the Tune icon in the System Setup window, or by right-clicking on a
sensor icon and selecting Tune. To save changes made during manual tuning,
press the Save Tune button on the Sensor Toolbar and the Tune will be saved
immediately. Other ways to save a Tune file include using File >> Save, File >>
Save As, or Tune >> Save Tune Parameters.