HAPSITE ER Operating Manual

3

As part of the HAPSITE’s preparation, it will run AutoTune. If AutoTune
successfully runs and completes, the AutoTune OK message will appear. Click
OK. See

Figure 5-3

. If AutoTune fails, see

Section 7.4, AutoTune Failure, on

page 7-6

. When the final results message is displayed, click OK.

Figure 5-3 AutoTune Complete

4

The HAPSITE will check pressures and automatically heat all necessary zones
to the setpoint temperatures specific to the selected method. Progress will be
indicated by a bar graph. Once all temperature zones have reached their
setpoints, a prompt will appear Press RUN to start method.

5

Click the RUN button on the pop-up window or from the control panel on the
screen. See

Figure 5-4

.

Figure 5-4 Run Button

6

Sample the background for one minute and note the TIC count.

7

Hold the probe over the sample of interest for up to 1 minute. If the TIC count
begins to rise, slowly back the probe away from the sample. If the TIC count
does not rise after a full minute, move the probe away from the sample.

7a

A peak may appear if the compound present is greater than 1 ppm. A
compound identification may also be present on the HAPSITE screen.

NOTE: No response may indicate either the compound present is less than the

detection limit, or that no detectable compound is present.