INFICON HAPSITE ER Chemical Identification System User Manual
Page 211
Advertising
5 - 17
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
10
Click RUN button on the pop-up window or from the Control Panel on the
screen.
Figure 5-22 Run Button
11
The HAPSITE will prompt Collecting Sample Now For. The Headspace will
automatically start sampling. No action is required by the user.
12
When the HAPSITE screen prompts Sampling is Done, the Headspace has
finished sampling. Again, no action is required by the user.
Figure 5-23 HSS Sampling Done
13
The chromatogram, showing the run’s progress, will appear on the Laptop
screen.
14
When prompted at the end of the run, place the Headspace needle into clean,
empty purge vial and click Select. See
Figure 5-24
.
Figure 5-24 Purge Vial Prompt