INFICON HAPSITE ER Chemical Identification System User Manual

Page 373

Advertising
background image

11 - 35

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

The SIM page for Survey mode provides the ability to create only one Scan Set of
multiple masses. Refer to

Figure 11-26

.

Scan Set Name . . . . . . . . . . . . . . . . A name should be created here to reference

the specific set of ions being detected.

Timed Mode. . . . . . . . . . . . . . . . . . . When selected, allows the time of the survey

method to be set to a determined length. This
method will run for a programmed amount of
time.

Trigger Mode. . . . . . . . . . . . . . . . . . When selected, programs this method to

start or stop based on the Run button.
Pressing the Run button turns on the sample
pump. This signals the beginning of the run.
Pressing Run again will turn off the sample
pump and end the sample run.

Scans to Average . . . . . . . . . . . . . . This determines how many scans will be

collected and averaged before the results are
updated on the chromatogram.

Advertising