HAPSITE ER Operating Manual
CAUTION
All caps must be in place and the front door must be
firmly shut before decontaminating the Analytical
Module.
WARNING
Make sure that you are in a well ventilated area.
WARNING
Wear appropriate PPE (Personal Protective Equipment).
Refer to the MSDS for the appropriate PPE.
12
Wipe the Analytical Module down using a soft cloth and a 10% hydrochlorite
(bleach) solution (i.e., 1 part bleach to 9 parts water).
13
Use a soft cloth and warm water and wipe the Analytical Module down again.
16.6 NEG Troubleshooting
If an MS Pressure Warning message is received on the front panel of the
HAPSITE, go to the TUNE STATUS option on the PROBES & SM page menu. If
seen on the Laptop, (see
) go to the Service Module Tab in the
Properties menu on the Laptop in ER IQ to check the MS Pressure. See
Figure 16-7 MS Pressure Error