HAPSITE ER Operating Manual

5

Replace the VI valve into the bottom of manifold. Hand tighten.
See

Figure 16-20

.

Figure 16-20 Replacing the VI Valve

6

Clean the o-ring on the Service Module using Methanol. See

Figure 16-21

.

Figure 16-21 Cleaning O-ring on the Service Module

7

Place the HAPSITE on the Service Module and lock down with the black side
latches. Refer to

section 2.6.2, Placing the HAPSITE on the Service Module,

on page 2-21

for additional instructions, if needed.

8

Continue with

section 16.7.4, Part 3: Install the New NEG Pump, on page

16-24

.