HAPSITE ER Operating Manual
5
Replace the VI valve into the bottom of manifold. Hand tighten.
See
Figure 16-20
.
Figure 16-20 Replacing the VI Valve
6
Clean the o-ring on the Service Module using Methanol. See
Figure 16-21
.
Figure 16-21 Cleaning O-ring on the Service Module
7
Place the HAPSITE on the Service Module and lock down with the black side
latches. Refer to
section 2.6.2, Placing the HAPSITE on the Service Module,
for additional instructions, if needed.
8
section 16.7.4, Part 3: Install the New NEG Pump, on page
.