HAPSITE ER Operating Manual
4
Remove the new NEG Pump from the protective container.
See
Figure 16-25
.
CAUTION
The NEG Pump is stored under a vacuum. The shelf life
is 5 years. If the vacuum is compromised, shelf life is
reduced to 1 year.
5
Remove the o-ring and seating ring assembly from the new NEG Pump.
See
Figure 16-25
.
Figure 16-25 New NEG Pump
6
On the new NEG Pump, wipe inside the flange with methanol.
See
Figure 16-26
.
Figure 16-26 Cleaning Flange with Methanol
O-ring and Seating
Assembly