HAPSITE ER Operating Manual

4

Remove the new NEG Pump from the protective container.
See

Figure 16-25

.

CAUTION

The NEG Pump is stored under a vacuum. The shelf life
is 5 years. If the vacuum is compromised, shelf life is
reduced to 1 year.

5

Remove the o-ring and seating ring assembly from the new NEG Pump.
See

Figure 16-25

.

Figure 16-25 New NEG Pump

6

On the new NEG Pump, wipe inside the flange with methanol.
See

Figure 16-26

.

Figure 16-26 Cleaning Flange with Methanol

O-ring and Seating

Assembly