7 - 15

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

7.5.3.1 Tune Parameters

EM Volt. . . . . . . . . . . . . . . . . . . . . . . Used to increase or decrease the gain of the

system. EM voltage should be set to a value
that achieves a Base Peak Gain of between
0.5 and 2.0.

Emission Current . . . . . . . . . . . . . . Used to optimize the ionization efficiency of

the ionizer. Emission Current is set to
achieve maximum intensity for mass 117.
Range is 100 - 1000. (350 is typical.)

Extractor Volt . . . . . . . . . . . . . . . . . Used to optimize the ionization efficiency of

the ionizer. Extractor Volt is set to achieve
maximum intensity for mass 117. Range is
0 - 100.

Focus Volt . . . . . . . . . . . . . . . . . . . . Used to optimize the ionization efficiency of

the ionizer. Focus Volt is set to achieve
maximum intensity for mass 117. Range is
-12 to +12.

Threshold . . . . . . . . . . . . . . . . . . . . One standard deviation of the Baseline.

Threshold determines if a measured point is
used in the peak area integration. If the point
is used, the baseline is subtracted before
use.

Baseline. . . . . . . . . . . . . . . . . . . . . . The mean value of the measured noise level.

Reverse Rod Polarity . . . . . . . . . . . Changes the Rod polarity on the Mass filter

and is selected for optimum performance at
mass 117.

TP. . . . . . . . . . . . . . . . . . . . . . . . . . . The total MS pressure.

Running Tune Base Peak Gain . . . Current measured Base Peak Gain (BPG).

NOTE: The Base Peak Gain will switch to

Red when BPG is outside the target
range.

Auto Resolve . . . . . . . . . . . . . . . . . Adjusts the resolution of all Mass Peaks to

the target resolution.

Save Tune . . . . . . . . . . . . . . . . . . . . Save the Tune File

Mass Cal.. . . . . . . . . . . . . . . . . . . . . Checks and corrects the ten calibration

masses for correct location within the mass
range.

Full Scan . . . . . . . . . . . . . . . . . . . . . Switches between Full Scan display mode

and Peak Scan display mode