7.4 AutoTune Failure

Occasionally, AutoTune may fail. Follow the procedures outlined below before
continuing.

1

When AutoTune fails, check Internal Standard canister pressure and expiration
date. Replace if necessary.

2

Rerun AutoTune by touching the HAPSITE icon on the front panel, touching the
HAPSITE System icon, touching the Tune icon and then, the Run AutoTune
button.

3

If AutoTune fails again, select File on the Laptop. From the drop down menu,
select View Tune Reports. See

Figure 7-9

.

Figure 7-9 Selecting View Tune Reports