process. If a NEG has greater than 50 hours (approximately 1/2 the expected life)
the more economical choice may be to replace the pump, based on the reduced
lifetime after exposure to atmospheric pressure.

Tools/equipment required

Flat head screw driver

Ionizer removal tool

13 mm open end wrench

13 mm torque wrench.

Methanol

Lint free cloth

Powder free gloves

Replacement aluminum seal ring (if NEG is to be installed)

CAUTION

If the NEG is to be reinstalled, the NEG Pump must be
removed when it is at room temperature. If the NEG Pump
has been used within 8 hours, let it cool overnight prior
to removal.

1

This procedure requires the use of a Service Module. The HAPSITE should be
physically on the Service Module and attached. Refer to

section 2.6, Service

Module, on page 2-17

and

Chapter 15, Service Module

for more information on

setting up and using the Service Module.

2

If the NEG Pump is hot from recent use, let the system cool overnight before
performing this procedure. Refer to

section 16.7.2, Part 1: NEG Pump

Removal, on page 16-19

.

WARNING

Do not bypass this cool down step!

3

Vent the HAPSITE by choosing Vent HAPSITE on the Service window of the
Service Module tab in the Properties window. This starts a process which
includes monitoring the cooling of the NEG Pump. See

Figure 16-43

.